KM940 Hand-held Combustion Analyser
- 10 - Once the time has reached zero an audible beep will be heard and will show the selected fuel on the following display:- NATURAL GAS ∗P
- 11 - 4.3 Main Displays The main display can be changed to show either 4 or 8 parameters at one time. Two options are available when 4 parame
- 12 - 4.3.2 Line Scroll Mode Line scroll mode allows you to customise the display. Use the and keys to change the bo
- 13 - 4.4 Sampling the Flue Gas Once the automatic calibration procedure has been completed and the specific fuel has been selected (See
- 14 - 4.6 Regular Checks During Sampling Care must be taken at all times not to exceed the analysers operating specifications, in particular
- 15 - 4.8 Electromagnetic Compatibility The European Council Directive 89/336/EEC requires that electronic equipment does not generate elect
- 16 - 5. MOVING THROUGH THE MENUS 5.1 Basic Operation From the MAIN DISPLAY NETT C . . . . 0.0 O2 % . . . 20.9
- 17 - 5.2 Menu Options and Settings 5.2.1 Main Menu The MAIN MENU consists of 4 sub menus which are shown below and detailed on the follow
- 18 - Calculation of fuel constants are detailed in the Appendix. Fuel constants will have to be calculated before a user fuel can be enter
- 19 - Selecting YES and will display the following screen. RESET SENSORS O2 % : 20.9 CO & NO = 0 PRESS ENTER MENU TO ESCAP
- 2 - CONTENTS Page No: 1. ANALYSER LAYOUT AND FEATURES... 4-7 1.1 Instrument
- 20 - 5.2.4 Display Menu LIGHT MODE CONTRAST : OFF : 8-PAGE : DEFAULT Allows the configuration of the display to be changed. LIGHT : Choo
- 21 - Once an alarm has been exceeded the display will flash every two minutes warning the user of an alarm state and display the gas concent
- 22 - Header : Allows two lines of 20 characters to be programmed into the analyser. The header appears on the top of the standard printout.
- 23 - 6. PRINTING INFORMATION Supplied as accessories for the KANE940 are an infra-red thermal printer or a dot matrix serial printer. Re
- 24 - 7. STORING AND RETRIEVING DATA The KANE940 can store up to 100 combustion tests. Once stored, the data can be viewed on the display o
- 25 - To print the data press . In the screen shown above locations 1 to 10 will be printed. During printing the following will b
- 26 - 8. MAINTENANCE 8.1 Emptying and Cleaning the In-line Water Trap The in-line water trap should be checked and emptied on a regular bas
- 27 - 9. PROBLEM SOLVING The following is a list of problems that may occur on the instrument through its operating life. If the cause of the
11. PRODUCT SPECIFICATION Parameter Resolution Accuracy Range Temp Measurement Flue Temperature with probe 1.0ºC/F +2.0ºC +0.3% reading 0-600º
- 29 - Dimensions Weight Handset Probe 1kg 220mm x 55mm x120mm L 420mm x Dia 8mm with stainless steel shaft, type K thermocouple and 3m hos
- 3 - 8. MAINTENANCE ...26 8.1 Emptyi
- 30 - APPENDICES A - Main Display Parameters The parameters and their meanings are detailed as follows : - DATE : Analyser date. See Set-U
- 31 - (Ta) temperature calculation if an INLET probe is not fitted. CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by cal
- 32 - B. COMBUSTION EFFICIENCY CALCULATION The efficiency calculation is based upon British Standard BS845. This identifies three sources o
- 33 - Calculated data: Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unburned carbon l
- 34 - C. CALCULATION OF FUEL DATA For any fuel not specified by Kane International the net calorific value, gross calorific value and compo
- 35 - D. ELECTROMAGNETIC COMPATABILITY (CE) STATEMENT This product has been tested for compliance with the following generic standards
- 4 - 1. ANALYSER LAYOUT AND FEATURES 1.1 Instrument Features and Keypad ON/OFF UP Scrolls up through
- 5 - 1.2 Instrument Layout (Rear) SENSORS FITTEDCharger SocketNOTE! - CoverDoes Not RemoveSerial NumberOptions LabelExhaust Port*NOTE! Do not
- 6 - 1.3 Standard Probe Configuration
- 7 - 1.4 Analyser Connections Gas Connectionto InstrumentParticle FilterIn-line Water TrapNeoprene HoseFlue TemperatureConnector
- 8 - 2. SAFETY WARNING This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted
- 9 - 4. NORMAL START UP SEQUENCE 4.1 Every Time You Use The Analyser BEFORE SWITCH-ON CHECK THAT: the particle filter is not dirty t